Getting Started HGCROC & Calibrations
This page describes the steps of starting up the HGCROC, programming it and running through the calibration procedures.
Programming KCU
First of all you need to open source the environment for vivado and open it (see commands on ORNL-DAQ computer below).
// sourcing vivado environment
source Xiling/2025.1/Vivado_Lab/setting64.sh
// open vivado lab
vivado_labThis should bring you to the home screen of vivado. There you should follow the steps listed below
Open the hardware manager
Switch on KCU (if not yet done)
Open target -> auto connect
program xckcu - firmware version for 1G & 1 HGCROC proto board (v2)
press play to check setup for each ASIC
Configuring the HGCROC
The software to configure the HGCROC is called H2GConfig (see link for details where to find it). This can be started as follows:
// starting HGCROC configuration program
python3 H2GConfig.py
This brings up the initial start window make sure you choose the appropriate number of KCU's and Asics (HGCROCs) per KCU. Which should bring you to the main window.

To test whether the KCUs are reachable use the Ping FPGA button, if that does not work check your network configuration and follow the steps laid out in link.
Configuration Files
In the config folder, you can find the configuration files for the different configurations. Several configurations are provided as examples.
udp_config.json: Configuration for the UDP communication. Including the IP addresses and ports for the different KCUs and the PC.h2gcroc_1v4_r1.json: Configuration for the H2GCROC registers. Users should not modify this file.Example_config.json: Example I2C configuration for testing the application.
After the first run, a config.ini file will be created in the same folder as the application. This file stores the last configuration used by the user. Deleting this file will reset the configuration to the default one.
For a brand new start of the setup please load the configurations
config_a0.jsonforasic 0config_a1.jsonforasic 1
You can either do this by pressing Send Current ASIC Config for a single asic or by Send All Config for all asics. After reconfiguring the asic using the configuration tool it should be checked that the the data lines and trigger lines have reasonable values using vivado for every asic (press play and check the values).

Manual modifications of Configuration values
The HGCROC has a very large number of configuration parameters and consequently registers which can and/or have to be set. A detailed description of those to the best current knowledge can be found in:
Nearly all of them can be adjusted using the H2GConfig gui in its various tabs which can be accessed via the side panel.

Below you find examples of each of the individual register panels, clicking on any of the bit fields will change the bit (0->1 or vice versa) so be careful when clicking randomly in the various positions.









In order to correctly position the maximum of the signal with respect to the trigger signal you might need to touch the L1_offset bits, they can be found under Digital_Half_0 & Digital_Half_1 remember to change them for all necessary ½ chips. The L1_offset is an 8bit number increasing the least significant bit by 1 will shift the waveform to the right by 1 sample.
HGCROC calibration procedure
For the HGCROC there are 4 main calibration steps:
IODelay scan: Although this isn't really a calibration you'll need to run it everytime you reconfigured the asic (otherwise you data might be nonesense)
Pedestal Calibration: Determines where the minimum
ADCvalue without any signal should be situated (recommended value80)ToA Calibration: Determines at which threshold value equivalent the Time of Arrival (ToA) is fired (rising edge). ATTENTION: The value we are setting here is given as
injection DAC(Digital-to-Analog-Converter) which is not equivalent to theADCvalue referred to during the pedestal calibration. The actual conversion factor between those depends on the SiPM type.ToT Calibration: Determines at which threshold value equivalent the Time over Threshold (ToT) is fired (falling edge). Once more the settable value is the
injection DACvalue no theADCvalue.


After these calibrations have been run the 2.5V Internal Injection test should be run ideally for all channels to validate the thresholds also in ADC equivalent. These test however take very long (~8min/4 channels) and it might not be advisable to run during a test beam setting.
The calibration is handled and steered using H2GCalib_3B package (link). Its workflow is as follows.

Calibration Gui
It is highly recommended to run the 100_UI.py script first. This script provides a user interface for selecting the calibration files and setting various parameters for the calibration process. Once more it will ask in the beginning to select how many KCU's and ASICs per KCU are connected.


Always start 101_SocketPool before running any other script. This script will forward the UDP packets from the H2GCROC-3B to the calibration scripts. It is essential for the calibration process to function correctly. Don't forget to terminate the socket before starting any other data taking process.
In the FPGA tab, you can do the IODelay setting, pedestal calibration, ToA calibration, and ToT calibration.
Config Folder: This is the json file that specifies the UDP address and port of the KCU system.
The current default configuration:
config/h2gcroc_1v4_r1.json
IO-delay scan
(Estimated running time: < 1 minute)
The IO Delay Scan is used to adjust the timing of the signals received by the FPGA. This is crucial for ensuring that the signals are correctly aligned and processed. Run this every time you do a power cycle / reprogram the FPGA.
-r Reset: If checked, the H2GCROC3B will be hard-reset before starting the scan.
-t Trigger: If checked, the trigger lines (T0, T1, T2, T3) will also be aligned.
-p Phase: This is the clock phase setting for the ASICs, make sure it matches the -i I2C JSON files in the following calibration steps.
You can find the results of the IO Delay Scan in the ./dump/102_IO_Delay_data_YYYYMMDD_HHMMSS folder. There will be pdf files for how the IO Delay values are set. The dashed red line indicates the optimal IO Delay value, which should be in the middle of the widest locked region.

After the IO-delay scan finished one should check in vivado that the data lines in case of a valid data flag are set to acccccc , while the trigger lines should be should show a different value.
Pedestal Calibration
(Estimated running time: ~ 3 minutes)
-t Pedestal Target [ADC]: This is the target pedestal value for all the channels. The script will adjust the pedestal values to achieve this target. Recommended value range: 50-150 ADC.
-i I2C JSON: This is the base I2C register configuration file for the H2GCROC-3B. The output register configuration will be based on this file.
The output of the pedestal calibration will be new I2C JSON files. And the file paths will be automatically updated into the -i field in the ToA calibration section. The output files will be saved in the ./dump/103_PedestalCalib_data_YYYYMMDD_HHMMSS folder. The result pdf file will show how the pedestal values are set.
As current default configuration config/default_2024Aug_config.json should be used. Make sure the fine channel calibration looks reasonable and no large outliers can be found. If the results aren't satisfactory reset the IO-delays and retry. If after a second attempt the calibration still hasn't succeeded. It could be tried to load a valid calibration from a different proto-board, primarily the Noinv_vref and Inv_vref bits in the Reference Voltage Register should be adjusted in that case.


A good calibration should have the final pedestal values nicely aligned around the target value, with a small spread across all channels.


Time-of-Arrivel(ToA) Calibration
(Estimated running time: ~ 45 minutes)
-t ToA Target [DAC]: (NOTE: it is by the unit of injection DAC, not ADC!) This is the target ToA value for all the channels. The script will adjust the ToA values to achieve this target. Recommended value range: 20-140 DAC.
-i I2C JSON: This is the base I2C register configuration file for the H2GCROC-3B. The output register configuration will be based on this file.
The output files will be updated in the -i field in the ToT calibration section. The output files will be saved in the ./dump/104_ToACalib_data_YYYYMMDD_HHMMSS folder.
The ToA-calibration is currently being executed in 5 steps, starting with a very coarse scanning window to estimate the first ToA_vRef values for the different half chips and then decreasing the stepping size an window size successively. The red line indicated in each vertical column indicated the edge at which the ToA starts firing. These need to be adjusted such that they correspond to the same value (or similar) value for all channels.

A good calibration should have the final ToA values like, however the timing alignment of the different halfs (color scale) is currently not performed.

Time-over-Threshold(ToT) Calibration
(Estimated running time: ~ 60 minutes)
-t ToT Target [DAC]: (NOTE: it is by the unit of injection DAC, not ADC!) This is the target ToT value for all the channels. The script will adjust the ToT values to achieve this target. Recommended value range: 200-700 DAC.
-i I2C JSON: This is the base I2C register configuration file for the H2GCROC3B. The output register configuration will be based on this file.
The output folder is ./dump/105_ToTCalib_data_YYYYMMDD_HHMMSS. The result pdf file is in the same format as the ToA calibration. Similarly as for the ToA calibration the calibration is run in 5 steps.

An example of a good ToT calibration result is shown below:

2.5V DAC Internal Injection
(Estimated running time: ~ 50 minutes for all channels and all phases)
This script is used to inject a 2.5V DAC signal into the H2GCROC3B channels for calibration purposes. It is not part of the main calibration flow but can be used for additional testing and verification.
DAC Value: This is the 12-bit DAC value to be injected into the channels.
Channel Start: the starting channel number for the injection.
Channel Count: the number of channels to be injected for each ASIC.
-i I2C JSON: This is the base I2C register configuration file for the H2GCROC3B. The output register configuration will be based on this file.
The output files will be saved in the ./dump/106_Injection2V5_data_YYYYMMDD_HHMMSS folder. The result pdf file will show the injected signal and the response of the channels. This can be used to verify the calibration results and the performance of the H2GCROC3B.



Loading Final calibration
After running the calibrations for all KCU's (FPGA) and HGCROCs (asics) move the files to a common location and name the appropriately. In case you are using an older configuration (i.e. preprepared) make sure the UDP settings atop the json file are correct, adjust them if necessary.
"UDP Settings": {
"IP Address": "10.1.2.208",
"Port": "11000"
},
"Target ASIC": {
"FPGA Address": 0,
"ASIC Address": 0
},Afterwards, load these using the H2GConfig for the respective asics.